HUANG Xuan,LI Fengning,ZHANG Fan.A cadaveric study on establishing individualized navigation templates for axis laminar screw insertion by using a three-dimensional printing forming technique[J].Chinese Journal of Spine and Spinal Cord,2014,(4):366-371.
A cadaveric study on establishing individualized navigation templates for axis laminar screw insertion by using a three-dimensional printing forming technique
Received:January 26, 2014  Revised:February 24, 2014
English Keywords:Laminar screw  Individualized navigation templates  Three-dimensional printing  Stereo lithography apparatus  Fused deposition modeling  Polylactic acid
Fund:国家自然科学基金资助项目(编号:81372012);长海医院“1255”学科建设计划资助项目(编号:CH125540200,CH125520900)
Author NameAffiliation
HUANG Xuan Department of Orthopedics, Changhai Hospital, Second Military Medical University, Shanghai, 200433, China 
LI Fengning 第二军医大学长海医院骨科 200434 上海市 
ZHANG Fan 第二军医大学长海医院骨科 200435 上海市 
王 琨  
陈 智  
阳青松  
党瑞山  
杨学东  
沈洪兴  
李 明  
Hits: 2580
Download times: 2344
English Abstract:
  【Abstract】 Objectives: To investigate the feasibility of establishing individualized navigation templates for axis laminar screw insertion by using a three-dimensional printing forming technique, and to evaluate the accuracy of template-assisted laminar screw insertion. Methods: Nineteen adult cervical specimens were selected to take CT-scan. After original Dicom data were imported into the Mimics software, the axis model was established to generate the cavity model based on a three-dimensional printing forming technique. After using a free-hand procedure to create a navigation template with a well-established laminar screw path, bilateral laminar screws were inserted assisted by the navigation templates. CT-scan was taken postoperatively to access the position of the screws. Results: Establishing axis and the related cavity models based on CT-scan images proved to be a feasible and precise method. Thirty-eight laminar screws were implanted assisted by 19 navigation templates. CT-scan images indicated all the screws were completely contained within the laminae. No screw perforated laminar walls or transverse foramen. Conclusions: Axis laminar screw insertion assisted by a three-dimensional printing model is highly accurate and simple which can be a new alternative to conventional technique.
View Full Text  View/Add Comment  Download reader
Close